Ventilation modes: | IPPV; A/C; PCV; SIMV; SIGH,Manual |
Ventilator parameter ranges | |
Graphical display: | |
P-T(pressure - time) | |
F-T(flow - time) | |
P-V loop (pressure - volume loop) | |
Alarm and protection | |
The AC power failure alarm | Power failure or no connection |
Internal battery backup low voltage alarm | <11.3±0.3V |
No tidal volume | ≤5Ml within 6s |
High oxygen concentration alarm Low oxygen concentration alarm | 19%-100% 18%-99% |
High Airway pressure alarm Low Airway pressure alarm High Minute Volume alarm Low Minute Volume alarm Continuous Pressure alarm | 20cmH2O-100cmH2O 0cmH2O-20cmH2O Adult(5L/min-20L/min) Paed(1L/min-15L/min 0-10L/min) (PEEP+1.5kPa) over 16s |
Suffocation warning | 5s-60s no spontaneous ventilation |
The maximum limited pressure | <12.5 kPa |
Fan error Oxygen deficit | Show on screen Show on screen |
Working conditions | |
Gas source | O2,N2O,AIR |
Pressure | 280kPa-600kPa |
Voltage | 100-240V |
Power frequency | 50/60Hz |
Input power | 85VA |
Vaporizer |
Anesthesia gas | Adjustable scope % (volume percentage) |
Halothane | 0 ~ 5 |
Enflurane | 0 ~ 5 |
Isoflurane | 0 ~ 5 |
Sevoflurane | 0 ~ 8 |
Trust point
Simplicity: 4 static casters with self-locking function.
Precision in an anesthesia ventilator with multiple ventilation modes: IPPV, A/C, PCV, SIMV,SIGH and MANUAL.
Flexible configurations able to customize your requirements.
International standard and advanced technology suitable for wide range use.
Compact interface and big screen give you better operating experience.
Over 5,000 units are installed in more than 200 countries.
Designed and manufactured by Superstar Medical with over 25 years of experience in ICU field.
Feature
10.4" TFT LCD screen displays the Ventilation parameters, Alarm information and Oscillogram.
High precision flowmeter, instantly know the fresh gas flow to your patient.
Integrated breathing circuit design, ensure easy operating and keep tidy.
Multiple working modes such as volume control and pressure limit, adapt to wide range patient.
Vaporizer with temperature, pressure, flow compensation and self-lock function, keep safety anytime.
Multiple parameters monitoring interface, make every parameter clear, let users know the patient conditions in all aspects.
Pressure-time, low-time loop oscillogram and high precision ETCO2, O2 concentration show in real time.
Vital sign monitor and anesthetic gas monitor are optional.
ETCO2 module and Anesthesia Gas Scavenging System (AGSS) are optional.
Safety
3 level alarm system, visual and sound alarm information.
With multiple type of alarm, reminder and protection functions.
Advanced power management control technology.
Built-in backup battery provide the emergency power supply to the unit.
Low O2 pressure alarm and N2O cut-off protection.